Patient Monitor HM11 - Parameter
Standard parameters DOWNLOAD FILE
ECG:
Lead selection: three/five lead standard; up to 7 leads can be displayed on the same screen
I, II, III, AVR, AVL, AVF, V chest leads
ECG detection range: 15~350bpm
Resolution: 1bpm
Measurement accuracy: the greater of ±2% or ±2bpm
Frequency Response: 0.67Hz-40 Hz
S-T segment monitoring: -2.0mV~2.0mV
Electrode drop indication: sound, light prompt
Scanning speed: 6.25, 12.5, 25, 50mm/sec
Gain selection: ×0.125, ×0.25, ×0.5, ×1, ×2, ×4, auto
ECG calibration: 1mV+5%
Protective isolation withstands 4000V/AC/50Hz voltage;
ECG cable lead wire: 3/5 lead
Standard configuration: adult universal ECG cable
Optional configuration: ECG cable for children and newborns
Alarm: The upper and lower limits of the alarm value can be set and automatically memorized; there is an alarm review;
Blood oxygen saturation:
Display: blood oxygen value, pulse histogram, waveform, pulse value
Blood oxygen range: 0%-100% adult/child/newborn
Resolution: 1%
Accuracy: ±3% (Undefined below 70%)
Pulse rate:
Measuring range: 30~280bpm
Pulse accuracy: ±2bpm
Alarm range: 20~300bpm, the lower limit cannot be greater than the upper limit
sensor:
Standard configuration: adult soft finger ring
Optional configuration: soft finger ring for children, newborn strap
Alarm value: upper and lower limits can be set, automatic memory
Non-invasive blood pressure:
Measurement method: Oscillation method
Measurement parameters: systolic blood pressure, diastolic blood pressure, mean blood pressure
Working mode: manual, automatic, continuous measurement
Unit: mmHg/kPa optional
Measurement interval time of automatic measurement mode: 2.5~120min, ten-level adjustable
Overvoltage protection: software and hardware overvoltage protection
Measuring range:
Adult Pediatric Newborn
Systolic blood pressure (mmHg) 40~255 40~200 40~135
Diastolic blood pressure (mmHg) 20~205 20~150 20~100
Average pressure (mmHg) 27~220 27~165 27~110
Cuff pressure range: 0-300 mmHg
Cuff:
Standard configuration: adult cuff
Select Configurations: Neonatal Cuff, Child Cuff
Alarm setting range:
Systolic blood pressure: 40~255mmHg, and the lower limit cannot be greater than the upper limit (adult),
40~200mmHg, and the lower limit cannot be greater than the upper limit (children),
40~135mmHg, and the lower limit cannot be greater than the upper limit (neonatal);
Diastolic blood pressure: 10~195mmHg, and the lower limit cannot be greater than the upper limit (adult),
10~150mmHg, and the lower limit cannot be greater than the upper limit (in children),
10~110mmHg, and the lower limit cannot be greater than the upper limit (neonatal);
Average pressure: 20~215mmHg, and the lower limit cannot be greater than the upper limit (adult),
20~165mmHg, and the lower limit cannot be greater than the upper limit (in children),
20~125mmHg, and the lower limit cannot be greater than the upper limit (neonatal);
Alarm display error: not more than ±5% of the set value.
Body temperature:
Body temperature measurement range: 0℃~50℃
Temperature measurement error: no more than ±0.1℃
Resolution: 0.1℃
Accuracy: ±0.1℃ (including sensor error)
Probe:
Standard configuration: permanent body temperature probe
Select configuration: esophageal temperature probe, rectal temperature probe.
Alarm value: upper and lower limits can be set, automatic memory
Alarm: All monitoring parameters can be set to alarm, and it has an alarm device that emits sound and light, and can cancel the alarm
Breathe:
Measurement method: thoracic impedance method
Respiratory rate measurement range and accuracy
Measuring range: adult: 0~ 120rpm;
Pediatric/Neonatal: 0~150rpm.
Measurement accuracy: 10~150 rpm ±2 rpm or ±2%, whichever is greater;
0~9 rpm is undefined.
Respiratory rate measurement resolution
Resolution: 1 rpm
Apnea alarm delay time
Can be set to: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
Respiratory rate alarm limit range
Adult pediatric/neonatal alarm limit can be set as: 2~150rpm
Please Note: Accessories such as trolleys need to be purchased together with the machine, and additional shipping will be charged for separate purchases (estimated between 300-600 US dollars)
Product Comparison Chart for ICU Patient Monitors
Configuration/
Parameters
|
HM12
US$1299
|
HM11
US$1099
|
HM10
US$799
|
Display Size
|
13.3-inch
|
12.1-inch
|
12.1-inch
|
Display Resolution
|
1366*768
|
1280*800
|
1024*768
|
Screen
|
Capacitive Touch Screen
|
Non touch screen,
Capacitive Touch Screen Optional
|
Non touch screen,
Resistive Touch Screen Optional
|
Multi-language
|
9
|
9
|
9
|
Lithium battery
|
Standard
|
Standard
|
Standard
|
Battery Life
|
7H
|
5H
|
5H
|
Screen Capture Function
|
Standard
|
Optional
|
Not Support
|
HDMI Output
|
Standard
|
Optional
|
Not Support
|
Parameter board
|
Fully discrete module
|
Six in one
|
Six in one
|
ECG Module
|
Independent
|
Integrated
|
Integrated
|
Blood Oxygen Module
|
Independent
|
Integrated
|
Integrated
|
Masimo SpO2
|
Optional
|
Optional
|
Not Support
|
Blood Pressure Module
|
Independent
|
Integrated
|
Integrated
|
Suntech Blood Pressure
|
Optional
|
Optional
|
Not Support
|
CO2 Mainstream
|
Support
|
Support
|
Support
|
CO2 By-Stream
|
Support
|
Support
|
Support
|
IBP Module
|
Optional
|
Not Support
|
Not Support
|
Nurse Call
|
Standard
|
Optional
|
Optional
|
Central Monitoring System
|
Optional
|
Optional
|
Optional
|
Internet Connection/LAN port
|
Support
|
Optional
|
Optional
|
ECG Leads
|
5 Leads 7 Leads
On The Same Screen
|
5 Leads 7 Leads
On The Same Screen
|
5 Leads 7 Leads
On The Same Screen
|
ECG Monitoring Range
|
15-350BPM
|
15-350BPM
|
15-350BPM
|
SpO2 Measurement Range
|
0%-100%
|
0%-100%
|
0%-100%
|
NIBP Measurement
|
0-300mmHg
|
0-300mmHg
|
0-300mmHg
|
History Review
|
1200H
|
1200H
|
960H
|
ECG Standards
|
AHA/IEC
|
AHA/IEC
|
AHA/IEC
|
Printing Module
|
Optional
|
Optional
|
Optional
|
WIFI
|
Optional
|
Optional
|
Optional
|
Net Weight
|
3.4Kg
|
3.25Kg
|
3.05Kg
|
Gross Weight
|
5.6Kg
|
5.4Kg
|
5.2Kg
|
Main unit Size
|
334±1Mm(L)*155±1Mm(D)*255±1Mm(H)
|
||
Overall Size
|
390±1Mm(L)*290±1Mm(D)*340±1Mm(H)
|