Patient monitor HM10 Technical Parameter
Technical parameter DOWNLOAD FILE
ECG:
Lead selection: three/five lead standard; up to 7 leads can be displayed on the same screen
I, II, III, AVR, AVL, AVF, V chest leads
ECG detection range: 15~350bpm
Resolution: 1bpm
Measurement accuracy: the greater of ±2% or ±2bpm
Frequency Response: 0.67Hz-40 Hz
S-T segment monitoring: -2.0mV~2.0mV
Electrode drop indication: sound, light prompt
Scanning speed: 6.25, 12.5, 25, 50mm/sec
Gain selection: ×0.125, ×0.25, ×0.5, ×1, ×2, ×4, auto
ECG calibration: 1mV+5%
Protective isolation withstands 4000V/AC/50Hz voltage;
ECG cable lead wire: 3/5 lead
Standard configuration: adult universal ECG cable
Optional configuration: ECG cable for pediatric and neonatal
Alarm: The upper and lower limits of the alarm value can be set and automatically memorized; there is an alarm review;
Blood oxygen saturation:
Display: blood oxygen value, pulse histogram, waveform, pulse value
Blood oxygen range: 0%-100% adult/child/neonatal
Resolution: 1%
Accuracy: ±3% (Undefined below 70%)
Pulse rate:
Measuring range: 30~280bpm
Pulse accuracy: ±2bpm
Alarm range: 20~300bpm, the lower limit cannot be greater than the upper limit
Sensor:
Standard configuration: adult soft finger ring
Optional configuration: soft finger ring for pediatric, neonatal strap
Alarm value: upper and lower limits can be set, automatic memory
Non-invasive blood pressure(NIBP):
Measurement method: Oscillation method
Measurement parameters: systolic blood pressure, diastolic blood pressure, mean blood pressure
Working mode: manual, automatic, continuous measurement
Unit: mmHg/kPa optional
Measurement interval time of automatic measurement mode: 2.5~120min, ten-level adjustable
Overvoltage protection: software and hardware overvoltage protection
Measuring range:
Item |
Adult |
Pediatric |
Neonatal |
Systolic blood pressure (mmHg) |
40~255 |
40~200 |
40~135 |
Diastolic blood pressure (mmHg) |
20~205 |
20~150 |
20~100 |
Average pressure (mmHg) |
27~220 |
27~165 |
27~110 |
Cuff pressure range: 0-300 mmHg
Cuff:
Standard configuration: adult cuff
Select Configurations: Neonatal Cuff, Child Cuff
Alarm setting range:
Systolic blood pressure: 40~255mmHg, and the lower limit cannot be greater than the upper limit (adult),
40~200mmHg, and the lower limit cannot be greater than the upper limit (pediatric),
40~135mmHg, and the lower limit cannot be greater than the upper limit (neonatal);
Diastolic blood pressure: 10~195mmHg, and the lower limit cannot be greater than the upper limit (adult),
10~150mmHg, and the lower limit cannot be greater than the upper limit (pediatric),
10~110mmHg, and the lower limit cannot be greater than the upper limit (neonatal);
Average pressure: 20~215mmHg, and the lower limit cannot be greater than the upper limit (adult),
20~165mmHg, and the lower limit cannot be greater than the upper limit (pediatric),
20~125mmHg, and the lower limit cannot be greater than the upper limit (neonatal);
Alarm display error: not more than ±5% of the set value.
Temperature:
Temperature measurement range: 0℃~50℃
Temperature measurement error: no more than ±0.1℃
Resolution: 0.1℃
Accuracy: ±0.2°C (including sensor error)
Probe:
Standard configuration: permanent body temperature probe
Select configuration: esophageal temperature probe, rectal temperature probe.
Alarm value: upper and lower limits can be set, automatic memory
Alarm: All monitoring parameters can be set to alarm, and it has an alarm device that emits sound and light, and can cancel the alarm
Breath measurement:
Measurement method: thoracic impedance method
Respiratory rate measurement range and accuracy
Measuring range: adult: 0~ 120rpm;
Pediatric/Neonatal: 0~150rpm.
Measurement accuracy: 10~150 rpm ±2 rpm or ±2%, whichever is greater;
0~9 rpm is undefined.
Respiratory rate measurement resolution
Resolution: 1 rpm
Apnea alarm delay time
Can be set to: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
Respiratory rate alarm limit range
Adult pediatric/neonatal alarm limit can be set as: 2~150rpm