Patient Monitor HM12 - Parameter

Technical parameter DOWNLOAD FILE

ECG:

Leads: 3/5 leads; most 7 ECG calculation leads can be displayed on the screen

Ⅰ、Ⅱ、Ⅲ、AVR、AVL、AVF、V leads

Measuring range:20~300bpm

Resolution:1bpm

Accuracy:the bigger one between ±2% or ±2bpm

Frequency response:0.67Hz-40 Hz

S-T segment monitoring:-2.0mV~2.0mV

Electrode drop indication:sound and light indicator

Sweep rate:6.25、12.5、25、50mm/sec

Gain :×0.125、×0.25、×0.5、×1、×2、×4、Auto

Electrical calibration:1mV+5%

Protection isolation withstands 4000V/AC/50Hz

Lead cable:3/5 lead

Standard configuration:universal adult ECG cable

Optional configuration:special ECG cable for Neonate and Pediatric

Alarm: upper and lower limits of alarm value can be set, automatic memory; alarm review

 

SpO2:

Display: SpO2 value, pulse column figure, waveform, pulse value

Measuring range: 0%-100% adult/pediatric/neonatal

Resolution: 1%

Precision: ±3% (less than 70% is not defined)

Pulse Rate:

     Measuring range:30~280bpm

Pulse precision:±2bpm

Alarm range: 20~300bpm, the lower limit can not be greater than the upper limit         

Sensor:

Standard configuration: adult finger clip

Optional configuration:pediatric finger clip, neonatal finger clip

Alarm value: the upper and lower limit can be set, automatic memory

 

NIBP:

Method: Oscillometric

Measuring parameter: Systolic pressure, Diastolic pressure, Mean pressure

Mode: Manual, Auto, STAT

Unit: optional for mmHg/ KPa

Measurement interval time of automatic measurement mode: 2.5~120min ten-level adjustable

Overpressure protection: software and hardware overpressure protection

Measuring range:

 

Adult

Pediatric

Neonatal

SYS(mmHg)

40~255

40~200

40~135

DIA(mmHg)

20~205

20~150

20~100

MEAN(mmHg)

27~220

27~165

27~110

 

Cuff pressure range: 0-300 mmHg

Cuff:    

Standard configuration: adult cuff

Optional configuration: neonatal cuff, pediatric cuff

Alarm setting range:

SYS:40~255mmHg, the lower limit can not be greater than the upper limit (adult),

        40~200mmHg, the lower limit can not be greater than the upper limit (pediatric),

        40~135mmHg, the lower limit can not be greater than the upper limit (neonatal);

DIA:10~195mmHg, the lower limit can not be greater than the upper limit (adult),

        10~150mmHg, the lower limit can not be greater than the upper limit (pediatric),

        10~110mmHg, the lower limit can not be greater than the upper limit (neonatal);

MEAN:20~215mmHg, the lower limit can not be greater than the upper limit (adult),

        20~165mmHg, the lower limit can not be greater than the upper limit (pediatric),

        20~125mmHg, the lower limit can not be greater than the upper limit (neonatal);

Alarm display error: Not greater than ±5% of the set value

 

Temperature:

Measuring Range:15℃~50℃

Measuring error: not more than ±0.1℃

Resolution:0.1℃

Accuracy:±0.2℃ (including sensor)

Probe:

Standard configuration: permanent body temperature probe

Optional configuration: Esophageal temperature probe, rectal temperature probe

Alarm value: the upper and lower limit can be set, automatic memory

Alarm: all monitoring parameters can be set for alarm, and it has a sound and light alarm device, and can cancel the alarm

 

RESP:

Method: the thorax impedance method

PR Measurement Range and Accuracy

Measurement range: Adult: 0~ 120rpm

                 Pediatric/Neonatal: 0~150rpm

Measurement Accuracy:10~150 rpm the bigger one between ±2 rpm or ±2%;

                    0~9 rpm not define.

PR measurement resolution

Resolution: 1 rpm

Suffocation alarm delay time

Set as:20s、25s、30s、35s、40s、45s、50s、55s、60s

PR alarm limit range

Adult neonatal/pediatric alarm limit can be set to: 2~150rpm

 

Product Comparison Chart for ICU Patient Monitors

Configuration/
Parameters
HM12
HM11
HM10
Display Size
13.3-inch
12.1-inch
12.1-inch
Display Resolution
1366*768
1280*800
1024*768
Screen
Capacitive Touch Screen
Non touch screen,
Capacitive Touch Screen Optional
Non touch screen,
 Resistive Touch Screen Optional
Multi-language
9
9
9
Lithium battery
Standard
Standard
Standard
Battery Life
7H
5H
5H
Screen Capture Function
Standard
Optional
Not Support
HDMI Output
Standard
Optional
Not Support
Parameter board
Fully discrete module
Six in one
Six in one
ECG Module
Independent
Integrated
Integrated
Blood Oxygen Module
Independent
Integrated
Integrated
Masimo SpO2
Optional
Optional
Not Support
Blood Pressure Module
Independent
Integrated
Integrated
Suntech Blood Pressure
Optional
Optional
Not Support
CO2 Mainstream
Support
Support
Support
CO2 By-Stream
Support
Support
Support
IBP Module
Optional
Not Support
Not Support
Nurse Call
Standard
Optional
Optional
Central Monitoring System
Optional
Optional
Optional
Internet Connection/LAN port
Support
Optional
Optional
ECG Leads
5 Leads 7 Leads
On The Same Screen
5 Leads 7 Leads
On The Same Screen
5 Leads 7 Leads
On The Same Screen
ECG Monitoring Range
15-350BPM
15-350BPM
15-350BPM
SpO2 Measurement Range
0%-100%
0%-100%
0%-100%
NIBP Measurement
0-300mmHg
0-300mmHg
0-300mmHg
History Review
1200H
1200H
960H
ECG Standards
AHA/IEC
AHA/IEC
AHA/IEC
Printing Module
Optional
Optional
Optional
WIFI
Optional
Optional
Optional
Net Weight
3.4Kg
3.25Kg
3.05Kg
Gross Weight
5.6Kg
5.4Kg
5.2Kg
Main unit Size
334±1Mm(L)*155±1Mm(D)*255±1Mm(H)
Overall Size
390±1Mm(L)*290±1Mm(D)*340±1Mm(H)