ICU CCU Ambulances Portable Vital Signs 6-Parameter Patient Monitor HM12 | DAWEI
ICU CCU Ambulances Portable Vital Signs 6-Parameter Patient Monitor HM12 | DAWEI
اتصل +8615305201757 ، إذا كنت بحاجة
HM12 Hospital ICU Patient Monitor - Performance

ICU CCU Portable Vital Signs 6-Parameter Medical Monitor
Standard configuration: ECG, SpO2, NIBP, RESP, Temperature, Heart rate;
Optional configuration: Printer, 3/5/12 leads ECG, masino SpO2, suntech NIBP, AG,C.O.,BIS, WIFI.
ECG:
Leads: 3/5 leads; most 7 ECG calculation leads can be displayed on the screen
Ⅰ、Ⅱ、Ⅲ、AVR、AVL、AVF、V leads
Measuring range:20~300bpm
Resolution:1bpm
Accuracy:the bigger one between ±2% or ±2bpm
Frequency response:0.67Hz-40 Hz
S-T segment monitoring:-2.0mV~2.0mV
Electrode drop indication:sound and light indicator
Sweep rate:6.25、12.5、25、50mm/sec
Gain :×0.125、×0.25、×0.5、×1、×2、×4、Auto
Electrical calibration:1mV+5%
Protection isolation withstands 4000V/AC/50Hz
Lead cable:3/5 lead
Standard configuration:universal adult ECG cable
Optional configuration:special ECG cable for Neonate and Pediatric
Alarm: upper and lower limits of alarm value can be set, automatic memory; alarm review
SpO2:
Display: SpO2 value, pulse column figure, waveform, pulse value
Measuring range: 0%-100% adult/pediatric/neonatal
Resolution: 1%
Precision: ±3% (less than 70% is not defined)
Pulse Rate:
Measuring range:30~280bpm
Pulse precision:±2bpm
Alarm range: 20~300bpm, the lower limit can not be greater than the upper limit
Sensor:
Standard configuration: adult finger clip
Optional configuration:pediatric finger clip, neonatal finger clip
Alarm value: the upper and lower limit can be set, automatic memory
NIBP:
Method: Oscillometric
Measuring parameter: Systolic pressure, Diastolic pressure, Mean pressure
Mode: Manual, Auto, STAT
Unit: optional for mmHg/ KPa
Measurement interval time of automatic measurement mode: 2.5~120min ten-level adjustable
Overpressure protection: software and hardware overpressure protection
Measuring range:
|
Adult |
Pediatric |
Neonatal |
SYS(mmHg) |
40~255 |
40~200 |
40~135 |
DIA(mmHg) |
20~205 |
20~150 |
20~100 |
MEAN(mmHg) |
27~220 |
27~165 |
27~110 |
Cuff pressure range: 0-300 mmHg
Cuff:
Standard configuration: adult cuff
Optional configuration: neonatal cuff, pediatric cuff
Alarm setting range:
SYS:40~255mmHg, the lower limit can not be greater than the upper limit (adult),
40~200mmHg, the lower limit can not be greater than the upper limit (pediatric),
40~135mmHg, the lower limit can not be greater than the upper limit (neonatal);
DIA:10~195mmHg, the lower limit can not be greater than the upper limit (adult),
10~150mmHg, the lower limit can not be greater than the upper limit (pediatric),
10~110mmHg, the lower limit can not be greater than the upper limit (neonatal);
MEAN:20~215mmHg, the lower limit can not be greater than the upper limit (adult),
20~165mmHg, the lower limit can not be greater than the upper limit (pediatric),
20~125mmHg, the lower limit can not be greater than the upper limit (neonatal);
Alarm display error: Not greater than ±5% of the set value
Temperature:
Measuring Range:15℃~50℃
Measuring error: not more than ±0.1℃
Resolution:0.1℃
Accuracy:±0.2℃ (including sensor)
Probe:
Standard configuration: permanent body temperature probe
Optional configuration: Esophageal temperature probe, rectal temperature probe
Alarm value: the upper and lower limit can be set, automatic memory
Alarm: all monitoring parameters can be set for alarm, and it has a sound and light alarm device, and can cancel the alarm
RESP:
Method: the thorax impedance method
PR Measurement Range and Accuracy
Measurement range: Adult: 0~ 120rpm
Pediatric/Neonatal: 0~150rpm
Measurement Accuracy:10~150 rpm the bigger one between ±2 rpm or ±2%;
0~9 rpm not define.
PR measurement resolution
Resolution: 1 rpm
Suffocation alarm delay time
Set as:20s、25s、30s、35s、40s、45s、50s、55s、60s
PR alarm limit range
Adult neonatal/pediatric alarm limit can be set to: 2~150rpm
Product Comparison Chart for ICU Patient Monitors
Configuration/ Parameters |
HM12 From US$1479 |
HM11 From US$1099 |
HM10 From US$869 |
Display Size | 13.3-inch | 12.1-inch | 12.1-inch |
Display Resolution | 1366*768 | 1280*800 | 1024*768 |
Capacitive Touch Screen | Standard | Optional | Optional Resistive Screen |
Battery Life | 7H | 5H | 5H |
Screen Capture Function | Standard | Optional | Not Support |
HDMI Output | Standard | Optional | Not Support |
Parameter board | Fully discrete module | Six in one | Six in one |
ECG Module | Independent | Integrated | Integrated |
Blood Oxygen Module | Independent | Integrated | Integrated |
Masimo SpO2 | Optional | Optional | Not Support |
Blood Pressure Module | Independent | Integrated | Integrated |
Suntech Blood Pressure | Optional | Optional | Not Support |
CO2 Mainstream | Support | Support | Support |
CO2 By-Stream | Support | Support | Support |
IBP Module | Optional | Not Support | Not Support |
Nurse Call | Standard | Optional | Optional |
Central Monitoring System | Optional | Optional | Optional |
Internet Connection/LAN port | Support | Optional | Optional |
ECG Leads | 5 Leads 7 Leads On The Same Screen |
5 Leads 7 Leads On The Same Screen |
5 Leads 7 Leads On The Same Screen |
ECG Monitoring Range | 15-350BPM | 15-350BPM | 15-350BPM |
SpO2 Measurement Range | 0%-100% | 0%-100% | 0%-100% |
NIBP Measurement | 0-300mmHg | 0-300mmHg | 0-300mmHg |
History Review | 1200H | 1200H | 960H |
ECG Standards | AHA/IEC | AHA/IEC | AHA/IEC |
Printing Module | Optional | Optional | Optional |
WIFI | Optional | Optional | Optional |
Net Weight | 3.4Kg | 3.25Kg | 3.05Kg |
Gross Weight | 5.6Kg | 5.4Kg | 5.2Kg |
Main unit Size | 334±1Mm(L)*155±1Mm(D)*255±1Mm(H) | ||
Overall Size | 390±1Mm(L)*290±1Mm(D)*340±1Mm(H) |
Please Note: Accessories such as trolleys need to be purchased together with the machine, and additional shipping will be charged for separate purchases (estimated between 300-600 US dollars)